DAQ Instrument

DAQ Instrument

Key Features

  • USB controlled

  • x16 Analog in, ±10V range

  • x2 Analog out, ±10V range

  • x8 Digital I/O, configurable logic levels

  • Frequency counter, 1 Hz - 50MHz

  • High-Z variant available

  • Internal reference clock

  • +5V/±15V/I2C/SPI expansion

  • Powered by +24V/+5V

  • Separate Power enable and reset

Description

The DAQ Instrument provides high-precision data acquisition and converts real-world signals to the digital domain for display, storage, and analysis. It has an onboard high-performance MCU with an Arm® Cortex®-M7 core running a RTOS (Real-Time Operating System), which also provides a USB FS/HS channel to an external host.

High-Z variant available bla bla

The instrument offers external connection to 16 single-ended analog input channels (AI) with a successive approximation ADC, two analog output channels (AO), 8 digital Push-Pull I/O channels (DIO). Each group of analog input channels AI0-7 and AO8-15 provides simultaneous sampling. All digital channels can be used as a counter. High-Z variant available where the signals AI0-15, AO0-1, DIO0-7 can be optically isolated.

Different types of external expansion such as number of analog and digital channels are possible by utilizing the built-in expansion interface.

DAQ Block Diagram

DAQ Instrument Block Diagram

DAQ Layout

DAQ Instrument Layout

Pinout

Pin

Name

Description

Pin

Name

Description

A1

+24V_IN

+24V Input
Power Supply
B1

USB_N

USB D- signal

A2

+24V_IN

+24V Input
Power Supply
B2

USB_P

USB D+ signal

A3

+24V_IN

+24V Input
Power Supply
B3

GND

Ground

A4

+24V_IN

+24V Input
Power Supply
B4

GND

Ground

A5

+5V_IN

+5V Input Power Supply

B5

GND

Ground

A6

+5V_IN

+5V Input Power Supply

B6

GND

Ground

A7

+5V_IN

+5V Input Power Supply

B7

GND

Ground

A8

+5V_IN

+5V Input Power Supply

B8

GND

Ground

A9

RST_N

Reset signal

B9

RSV

Reserved

A10

PWR_EN

Power enable of the instrument

B10

RSV

Reserved

A11

RSV

Reserved

B11

RSV

Reserved

A12

EXP_I2C_SCL/EXP_SPI_CS1

Expansion Interface I2C SCL or SPI CS1

B12

EXP_SPI_SCLK

Expansion Interface SPI Clock-signal

A13

EXP_I2C_SDA/EXP_SPI_CS2

Expansion Interface I2C SDA or SPI CS2

B13

EXP_SPI_MOSI

Expansion Interface SPI MOSI

A14

EXP_SPI_CS3

Expansion Interface UART TX or SPI CS3

B14

EXP_SPI_MISO

Expansion Interface SPI MISO

A15

EXP_SPI_CS4

Expansion Interface UART RX or SPI CS4

B15

EXP_SPI_CS0

Expansion Interface SPI CS0

A16

EXP_RST_N

Expansion Interface Reset output active low

B16

GND

Ground

A17

DIO0

Digital In/Out 0

B17

DIO1

Digital In/Out 1

A18

DIO2

Digital In/Out 2

B18

DIO3

Digital In/Out 3

A19

DIO4

Digital In/Out 4

B19

DIO5

Digital In/Out 5

A20

DIO6

Digital In/Out 6

B20

DIO7

Digital In/Out 7

A21

+5V_EXP

+5VDC Power Supply for Expansion

B21

GND

Ground

A22

+15V_EXP

+15VDC Power Supply for Expansion

B22

-15V_EXP

-15VDC Power Supply for Expansion

A23

GND

Ground

B23

GND

Ground

A24

AO0

Analog Output 0

B24

AO1

Analog Output 1

A25

AI0

Analog Input 0

B25

AI1

Analog Input 1

A26

AI2

Analog Input 2

B26

AI3

Analog Input 3

A27

AI4

Analog Input 4

B27

AI5

Analog Input 5

A28

AI6

Analog Input 6

B28

AI7

Analog Input 7

A29

AI8

Analog Input 8

B29

AI9

Analog Input 9

A30

AI10

Analog Input 10

B30

AI11

Analog Input 11

A21

AI12

Analog Input 12

B31

AI13

Analog Input 13

A32

AI14

Analog Input 14

B32

AI15

Analog Input 15

Absolute Maximum Ratings

Exceeding ABSOLUTE MAXIMUM RATINGS can cause permanent damage to the device.

Min

Max

USB_P, USB_N

-0.5 V

+5.5 V

+24V Input Power Supply

-0.3 V

+25.5 V

+5V Input Power Supply

-0.3 V

+5.2 V

Expansion +5V Output Power Supply Current

+200 mA

Expansion +15V Output Power Supply Current

+75 mA

Expansion -15V Output Power Supply Current

-75 mA

Voltage on any Expansion Digital Pin

-0.3 V

+3.6 V

Voltage on any Digital I/O Pin

-0.3 V

+5.5 V

Voltage on any Analog Input Pins

-10.5 V

+10.5 V

Typical Performance Characteristics

Analog Input Current vs. Input Voltage

DAQ Analog Input Current vs. Input Voltage (High-Z disabled)

Environmental

The DAQ Instrument is designed for following environmental conditions and for indoor use only.

Min

Max

Operating Temperature

0°C

+55°C

Operating Humidity`

10% RH

90% RH

Storage Temperature

0 °C

+70 °C

Storage humidity

5% RH

95% RH

Pollution Degree

2

Safety and Compliance

Safety

Update This Product is designed to meet the requirements of Electrical Safety Partial Testing according to EN 61010-1:2010 + A1:2019 + AC: 2019 and EN 61010-2-030:2021 + A11:2021.

Electromagnetic Compatibility

This product is designed to meet the requirements of EMC Testing according to EN 61326-1:2021 and CE standards of EMC for electrical equipment for measurement, control, and laboratory use.

CE Compliance

This product meets the essential requirements of applicable European Directives, as follows:

  • 2014/35/EU; Low-Voltage Directive (safety)

  • 2014/30/EU; Electromagnetic Compatibility Directive (EMC)

RoHS

Waste Electrical and Electronic Equipment (WEEE)

For EU Customers: at the end of the product life cycle, all Mikrodust products must be disposed of according to local laws and regulations.

Ordering information

Item

Part Number

AtoMik® Data Acquisition (DAQ) Instrument

IAD-1000-D1

AtoMik® Data Acquisition (DAQ) Instrument (High-Z)

IAD-1001-D1